Chapter 7 __________________________________________________ Getting the Data Messages

VAISALA______________________________________________________________________ 103

The structure of the supervisor response message:

M

=

Transducer id 1 units (hits/cm

2

, Hail peak intensity)

3

=

Transducer id for Hail peak intensity

*

Checksum delimiter

51

=

Two-character checksum for the response

<cr><lf>

=

Response terminator

where
$

=

Start of the message

WI

=

Device type (WI = weather instrument)

XDR

=

Transducer measurement response identifier

C

=

Transducer id 2 type (temperature), see the following
Transducer table

20.4

=

Transducer id 2 data (Heating temperature)

C

=

Transducer id 2 units (C, Heating temperature)

2

=

Transducer id for Heating temperature

U

=

Transducer id 0 type (voltage)

12.0

=

Transducer id 0 data (Heating voltage)

M

=

Transducer id 0 units (N = heating disabled or
heating temperature too high

1

, Heating voltage)

0

=

Transducer id for Heating voltage

U

=

Transducer id 1 type (Supply voltage)

12.5

=

Transducer id 1 data (Supply voltage)

V

=

Transducer id 1 units (V, Supply voltage)

1

=

Transducer id for Supply voltage

U

=

Transducer id 2 type (voltage)

3.460

=

Transducer id 2 data (3.5V reference voltage)

V

=

Transducer id 2 units (V, 3.5V reference voltage)

2

=

Transducer id for 3.5V reference voltage

G

=

Transducer id 4 type (generic)

HEL/___

=

Transducer id 4 data (info field)
Transducer id 4 units (none, null)