Chapter 8 ____________________________________________ Sensor and Data Message Settings

VAISALA______________________________________________________________________ 115

Changing the Settings (aTU)

You can change the following settings:

-

parameters included in the data message,

-

update interval,

-

pressure unit, and

-

temperature unit.

Make the desired setting with the following command. Select the
correct value/letter for the setting fields, see

Setting Fields on page 114

.

See the examples.

Command format in ASCII and NMEA 0183:

aTU,R=x,I=x,P=x,T=x<cr><lf>

Command format in SDI-12:

aXTU,R=x,I=x,P=x,T=x!

Examples (ASCII and NMEA 0183, device address 0):

You need the temperature and humidity data to be available in every 30
seconds

Changing the parameter selection:

0TU,R=0101000001010000<cr><lf>

0TU,R=01010000&01010000<cr><lf>

where
R, I, P, T

=

The pressure, temperature and humidity sensor
setting fields, see

Setting Fields on page 114

.

x

Value for the setting

<cr><lf>

=

Command terminator in ASCII and NMEA 0183

!

=

Command terminator in SDI-12

NOTE

Character '&' is not allowed in the command.