Setting fields – Vaisala WXT520 User Manual

Page 126

Advertising
background image

User’s Guide ______________________________________________________________________

124 _________________________________________________________________ M210906EN-C

Setting Fields

[R]

=

Parameter selection: This field consists of 16 bits
defining the supervisor parameters included in the
data messages. The bit value 0 disables and the bit
value 1 enables the parameter.

The bits 1-8 determine the
parameters included in the
message obtained with the
following commands:
-ASCII: aR5 and ar5
-NMEA 0183: $--WIQ,XDR*hh
-SDI-12: aM5, aMC5, aC5, and
aCC5
-SDI-12 continuous: aR5 and
aRC5

1st bit (most left)

Th Heating temperature

2nd bit

Vh Heating voltage

3rd bit

Vs Supply voltage

4th bit

Vr 3.5 V reference voltage

5th bit

Id Information field

6th bit

spare

7th bit

spare

8th bit

spare

&

delimiter

The bits 9-16 determine the
supervisor parameters
included in the composite data
message obtained with the
following commands:
-ASCII: aR0 and ar0
-NMEA 0183: aR0, ar0
-SDI-12: aM, aMC, aC, and aCC
-SDI-12 continuous: aR and
aRC

9th bit

Th Heating temperature

10th bit

Vh Heating voltage

11th bit

Vs Supply voltage

12th bit

Vr 3.5 V reference voltage

13th bit

Id Information field

14th bit

spare

15th bit

spare

16th bit (most right)

spare

[I]

=

Update interval: 1 ... 3600 seconds. When the
heating is enabled the update interval is forced to 15
seconds.

[S]

=

Error messaging: Y = enabled, N = disabled

[H]

=

Heating control enable: Y = enabled, N = disabled
Heating enabled: The control between full and half
heating power is on as described in

Heating

(Optional) on page 29

.

Heating disabled: Heating is off in all conditions.

<cr><lf>

=