Chapter 11 technical specifications, Performance, Chapter 11 – Vaisala WXT520 User Manual

Page 141: Table 12, Barometric pressure, Table 13, Air temperature, Technical specifications

Advertising
background image

Chapter 11 ____________________________________________________ Technical Specifications

VAISALA______________________________________________________________________ 139

CHAPTER 11

TECHNICAL SPECIFICATIONS

This chapter provides the technical data of Weather Transmitter
WXT520.

Performance

Table 12

Barometric Pressure

Property

Description/Value

Range

600 ... 1100 hPa

Accuracy

±0.5 hPa at 0 ... +30 °C (+32 ... +86 °F)
±1 hPa at -52 ... +60 °C (-60 ... +140 °F)

Output resolution

0.1 hPa, 10 Pa, 0.001 bar, 0.1 mmHg,
0.01 inHg

Units available

hPa, Pa, bar, mmHg, inHg

Table 13

Air Temperature

Property

Description/Value

Range

-52 ... +60 °C (-60 ... +140 °F)

Accuracy (for sensor element)
at +20 °C (+68 °F)
For accuracy over temperature
range, see the following graph

±0.3 °C

Output resolution

0.1 °C (0.1 °F)

Units available

°C, °F