Chapter 7 __________________________________________________ Getting the Data Messages

VAISALA_______________________________________________________________________ 89

Example 2:

Start a concurrent pressure, humidity and temperature measurement and
request the data:

0C2!000503<cr><lf> (measurement ready in 5 seconds and 3
parameters available, for aC command device address not sent as a sign
of a completed measurement)

0D0!0+23.6+29.5+1009.5<cr><lf>

Example 3:

Start a precipitation measurement and request the data:

0M3!00006<cr><lf> (6 parameters available immediately, thus the
device address is not sent)

0D0!0+0.15+20+0.0+0.0+0+0.0<cr><lf>

Example 4:

Start a supervisor measurement with CRC and request the data:

0MC5!00014<cr><lf> (measurement ready in one second and 4
parameters available)

0<cr><lf> (measurement completed)

0D0!0+34.3+10.5+10.7+3.366DpD<cr><lf>

Example 5:

Start a composite measurement and request the data. The configuration
of the parameter set is such that nine (9) parameters are available. Thus
start measurement command aM can be used. Due to the 35-character
limit in response message, aD0 returns only six parameters. The
remaining parameters are retrieved with aD1.

0M!00059<cr><lf> (measurement ready in 5 seconds and 9
parameters available)

0<cr><lf> (measurement completed)

0D0!0+340+0.1+23.7+27.9+1009.3+0.15<cr><lf>

0D1!0+0.0+0+0.0<cr><lf>