Chapter 8 ____________________________________________ Sensor and Data Message Settings

VAISALA______________________________________________________________________ 109

Setting Fields

The parameter order is shown in the following table:

[R]

=

Parameter selection: This field consists of 16 bits
defining the wind parameters included in the data
messages. The bit value 0 disables and the bit value 1
enables the parameter.

The bits 1-8 determine the
parameters in the data
message obtained with the
following commands:
-ASCII: aR1 and ar1
-NMEA 0183: $--WIQ,XDR*hh
-SDI-12: aM1, aMC1, aC1, and
aCC1
-SDI-12 continuous: aR1 and
aRC1

1st bit (most left)

Dn Direction minimum

2nd bit

Dm Direction average

3rd bit

Dx Direction maximum

4th bit

Sn Speed minimum

5th bit

Sm Speed average

6th bit

Sx Speed maximum

7th bit

spare

8th bit

spare

&

delimiter

The bits 9-16 determine the
wind parameters in the
composite data message
obtained with the following
commands:
-ASCII: aR0, ar0
-NMEA 0183: aR0, ar0
-SDI-12: aM, aMC, aC, and aCC
-SDI-12 continuous: aR and
aRC

9th bit

Dn Wind direction minimum

10th bit

Dm Wind direction average

11th bit

Dx Wind direction maximum

12th bit

Sn Speed minimum

13th bit

Sm Speed average

14th bit

Sx Speed maximum

15th bit

spare

16th bit (most right)

spare

[I]

=

Update interval: 1 ... 3600 seconds

[A]

=

Averaging time: 1 ... 3600 seconds
Defines the period over which the wind speed and
direction averaging is calculated. Same period is also
used for maximum and minimum calculation. See
also Appendix D

Wind Measurement Averaging

Method on page 163

for difference in averaging

practices when A<I and A>I.