Minimum reference mass – RADWAG HTY 60/H5/K Multifunctional Scales User Manual

Page 103

Advertising
background image

103

Options:

- Automatic correction of reference mass disabled

- Automatic correction of reference mass enabled

Function <

Automatic correction of reference mass>

in mode

<

Parts counting >

is enabled at the moment of estimating the sample

quantity and signalled by displaying <PCS> and <SMP> (single part/piece

mass) on the top part of the display.

There are four criteris of working “Automatic correction of reference
mass”

function:

1. equilibrium should be reached,

2. quantity of parts/pieces should be increased,

3. added quantity of parts/pieces should not be greater than double

number of parts/pieces on the pan,

4. the new sample can be different from the old sample by ± 0.3

of pcs (absolute value),

If a user recognises that the sample quantity is adequate the unit mass

(single piece mass) can be saved (see ch. 26.6 of this manual)

and disable the function by pressing ENTER/PRINT.

Notice:
While the parts counting function is active the functionality of
ENTER/PRINT
button is changed. Pressing
ENTER/PRINT does not result in printing and
saving weighings.

26.2.2. Minimum reference mass

It concerns the <

Parts counting > working mode

Users before beginning the procedure of single piese mass evaluation can

declare “minimum reference mass” i.e. minimum total weight of all pieces

put on the scale pan expressed in reading divisions.


Procedure:

Enter parameter group <

Working modes>

according to ch.

24 of this manual,