Installing in pue hy housing – RADWAG HTY 60/H5/K Multifunctional Scales User Manual
Page 219

219
platforms connected
to indicators in metal
housing via a cable
only
platforms electrically
connected to indicators’
metal housings e.g.
pillars, racks
Load cells without internal
shield connection to the
tensometer body
POINT B
E
Load cells with internal
shield connection to the
tensometer body
POINT B
POINT B
Point B
– screwed terminal electrically connected to the metal housing
of the indicator (possible using of soldering eye).
E
– soldering pad on a DP2 PCB.
The way of checking connection between the shield and the tensometer
body
Use an ohmmeter for this purpose.
37.5.4. Installing in PUE HY housing
DP2
modules are intended to mount inside PUE HY housings. It is mounted
to the main board to the 10-pin J11 connector. For DP2 module an additional
gland is installed on the back wall of the housing.
Installing procedure:
1. Unplug the scale from mains;
2. Unscrew and take off the back wall of the housing;
- HTY 60/H4/K Multifunctional Scales HTY 60/C2/K Multifunctional Scales HTY 6/H2/K Multifunctional Scales HTY 6/F1/K Multifunctional Scales HTY 300/H6/K Multifunctional Scales HTY 300/H5/K Multifunctional Scales HTY 300/C3/K Multifunctional Scales HTY 300/C2/K Multifunctional Scales HTY 30/H3/K Multifunctional Scales HTY 30/F1/K Multifunctional Scales HTY 30/C2/K Multifunctional Scales HTY 3/H1/K Multifunctional Scales HTY 150/H6/K Multifunctional Scales HTY 150/H5/K Multifunctional Scales HTY 150/H4/K Multifunctional Scales HTY 150/H3/K Multifunctional Scales HTY 150/C3/K Multifunctional Scales HTY 150/C2/K Multifunctional Scales HTY 15/H3/K Multifunctional Scales HTY 15/H2/K Multifunctional Scales HTY 15/F1/K Multifunctional Scales HTY 60/H3/K Multifunctional Scales