Send the stable result in basic unit, Send the result immediately in basic unit, Notice: use dots as decimal points in tare values – RADWAG HTY 60/H5/K Multifunctional Scales User Manual
Page 188: Syntax: s cr lf, Possible answers, Frame format, Syntax: si cr lf

188
Notice:
Use dots as decimal points in tare values.
34.4.5. Send the stable result in basic unit
Syntax: S CR LF
Possible answers:
S_A CR LF
S_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
S_I CR LF
- command comprehended but cannot be executed
S_A CR LF
MASS FRAME
- command accepted and in progress
- mass value in basic unit is returned
Frame format:
1
2-3
4
5
6
7-15
16
17
18
19
20
21
S
space
stability
space
sign
mass
space
unit
CR
LF
Example:
S CR LF
– computer command
S _ A CR LF
- command accepted and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF
– command done, mass value
in basic unit is returned.
34.4.6. Send the result immediately in basic unit
Syntax: SI CR LF
Possible answers:
SI_I CR LF
- command comprehended but cannot be executed at the moment
MASS FRAME
- mass value in basic unit is returned
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
I
space
stability
space
sign
mass
space
unit
CR
LF
- HTY 60/H4/K Multifunctional Scales HTY 60/C2/K Multifunctional Scales HTY 6/H2/K Multifunctional Scales HTY 6/F1/K Multifunctional Scales HTY 300/H6/K Multifunctional Scales HTY 300/H5/K Multifunctional Scales HTY 300/C3/K Multifunctional Scales HTY 300/C2/K Multifunctional Scales HTY 30/H3/K Multifunctional Scales HTY 30/F1/K Multifunctional Scales HTY 30/C2/K Multifunctional Scales HTY 3/H1/K Multifunctional Scales HTY 150/H6/K Multifunctional Scales HTY 150/H5/K Multifunctional Scales HTY 150/H4/K Multifunctional Scales HTY 150/H3/K Multifunctional Scales HTY 150/C3/K Multifunctional Scales HTY 150/C2/K Multifunctional Scales HTY 15/H3/K Multifunctional Scales HTY 15/H2/K Multifunctional Scales HTY 15/F1/K Multifunctional Scales HTY 60/H3/K Multifunctional Scales