Switch off continuous transmission in basic unit, Switch on continuous transmission in current unit, Switch off continuous transmission in current unit – RADWAG HTY 60/H5/K Multifunctional Scales User Manual
Page 191: Frame format, Syntax: c0 cr lf, Possible answers answers, Syntax: cu1 cr lf, Possible answers, Syntax: cu0 cr lf

191
C1_I CR LF
- command comprehended but cannot be executed
C1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in basic unit is returned
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
I
space
stability
space
sign
mass
space
unit
CR
LF
34.4.11. Switch off continuous transmission in basic unit
Syntax: C0 CR LF
Possible answers answers:
C0_I CR LF
- command comprehended but cannot be executed
C0_A CR LF
- command comprehended and executed
34.4.12. Switch on continuous transmission in current unit
Syntax: CU1 CR LF
Possible answers:
CU1_I CR LF
- command comprehended but cannot be executed
CU1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in current unit is returned
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
U
I
stability
space
sign
mass
space
unit
CR
LF
34.4.13. Switch off continuous transmission in current unit
Syntax: CU0 CR LF
Possible answers:
- HTY 60/H4/K Multifunctional Scales HTY 60/C2/K Multifunctional Scales HTY 6/H2/K Multifunctional Scales HTY 6/F1/K Multifunctional Scales HTY 300/H6/K Multifunctional Scales HTY 300/H5/K Multifunctional Scales HTY 300/C3/K Multifunctional Scales HTY 300/C2/K Multifunctional Scales HTY 30/H3/K Multifunctional Scales HTY 30/F1/K Multifunctional Scales HTY 30/C2/K Multifunctional Scales HTY 3/H1/K Multifunctional Scales HTY 150/H6/K Multifunctional Scales HTY 150/H5/K Multifunctional Scales HTY 150/H4/K Multifunctional Scales HTY 150/H3/K Multifunctional Scales HTY 150/C3/K Multifunctional Scales HTY 150/C2/K Multifunctional Scales HTY 15/H3/K Multifunctional Scales HTY 15/H2/K Multifunctional Scales HTY 15/F1/K Multifunctional Scales HTY 60/H3/K Multifunctional Scales