User adjustment, Adjusting procedure – RADWAG HTY 60/H5/K Multifunctional Scales User Manual
Page 84

84
22. USER ADJUSTMENT
An option only for non-verified scale
Scales require to recalculate internal divisions to more suitable ones (e.g.
g, kg etc.). In order to do this they require an adjustment factor. It is adjusted
during the adjustment procedure using a mass standard. Adjustment should
be carried out if weighing a standard mass shows a different mass value.
To enter < User Adjustment>,
press
key and then
:
„
User Adjustment”
.
22.1. Adjusting procedure
•
Enter submenu < User Adjustment > according to ch. 22
and select: “
Adjustment”
,
•
After entering the parameter the following message box appears:
•
Take the load off the pan of platform 1,
•
Press button
. The following message appears during
adjusting start mass: „Evaluation of starting mass”,
•
After the procedure has been completed the following message
box appears:
- HTY 60/H4/K Multifunctional Scales HTY 60/C2/K Multifunctional Scales HTY 6/H2/K Multifunctional Scales HTY 6/F1/K Multifunctional Scales HTY 300/H6/K Multifunctional Scales HTY 300/H5/K Multifunctional Scales HTY 300/C3/K Multifunctional Scales HTY 300/C2/K Multifunctional Scales HTY 30/H3/K Multifunctional Scales HTY 30/F1/K Multifunctional Scales HTY 30/C2/K Multifunctional Scales HTY 3/H1/K Multifunctional Scales HTY 150/H6/K Multifunctional Scales HTY 150/H5/K Multifunctional Scales HTY 150/H4/K Multifunctional Scales HTY 150/H3/K Multifunctional Scales HTY 150/C3/K Multifunctional Scales HTY 150/C2/K Multifunctional Scales HTY 15/H3/K Multifunctional Scales HTY 15/H2/K Multifunctional Scales HTY 15/F1/K Multifunctional Scales HTY 60/H3/K Multifunctional Scales