Commissioning, Error codes – IKA C-MAG HS 10 digital User Manual

Page 12

Advertising
background image

­­12

CMAG dig 0511

Commis-
sioning

Put device switch (A) in the OFF position

Plug in (G) mains cable

Once connected to the power supply the device
is in “stand-by” mode

Put device switch (A) in the ON position

Set the engine speed with the operating button (MOT)

Any set values are retained when device is switched
off and even after the device is disconnected from the
power!

Commissioning

Stirring

Heating

without external
temperature sensor

not in
operating mode A

with external
temperature sensor

(direct
temperature
control in the
medium)

Put device switch (A) in the OFF position

Plug in the PT1000 sensor into the socket (I)

Put device switch (A) in the ON position

Adjust the set-point temerature with the operating but-
ton (TEMP)

• LED (F) lights up
• the instantaneous temperature (K) is indicated on the
display (corresponds sensor- and/or medium temperature)
• indicated set point temperature (L) corresponds the
desired medium temperature
• LED (E) lights up and the symbol for energy input (M) is
indicated on the display (with the heating is switched on)
• indicate HOT (D) on the display with over 50°C on the
heating plate (in the agitating and stand-by-mode)

Put device switch (A) in the ON position

Set the operating mode (see capture “Operating
modes”)

Adjust the set-point temperature with the operating but-
ton (TEMP) (only be set in operating mode B)

• indicate the adjusted set point temperature on the

LCD display (D) position (L)

• indicated the set point temperature corresponds the

heating plate temperature

• LED (E) lights up and the symbol for energy input

(M) is indicated on the display (with the heating is
switched on)

• indicate HOT (D) on the display with over 50°C on

the heating plate (in the agitating and stand-by-
mode)

Any malfunctions during operation will be identified by an error mes-
sage on the display (D).

Proceed as follows in such cases:

Switch off device using the main switch (A)

Carry out corrective measures

Restart device

Error codes

Error

Cause

Solution

code

E1

Sensor not

• Connect sensor

connected

• Change operating mode

(operating mode A)

E3

Temperature inside • Switch off device and allow
device too high

to cool down

E5

Sensor not in

• Sensor in medium

media

immerse

E6

Sensor defectively • Exchange the sensor

E7

Hi-Temp

• Medium to cool down

temperature border
exceeded

E9

Initializations error

• Switch off/on the device

E11

Motor blocks

• use other stirring bar
• reduce viscosity

If the actions described fail to resolve the fault or another error
code is displayed then take one of the following steps:

- Contact the service department
- Send the device for repair, including a short description of the

fault.