11 Specifications

11.1 Power Supply

- Input voltage:

10.5 ... 35.0 V DC

- Residual ripple:

≤ 5 %

- Current consumption without sensors and outputs:

Typical (mean value):

≤ 90 mA Stand-By / ≤ 200 mA with radio connection

Pulsed:

≤ 2 A (power supply recommendation: 3.5 A)

- Reverse voltage protection:

-35 V

11.2 Sensor Inputs

(5 pole female connection "Sensor 1", 8 pole female connection "Sensor 2")

A maximum of 2 SMART sensors can be connected.

- Output voltage (+U

B out

): +U

B in

– 0.5V

- Current supply:

max. 500mA at 50°C

11.3 Logic measurement channels

A total of up to 32 measurement channels can be processed by the CSI-F-10.
A measurement channel can be a sub-channel of a SMART sensor or a value derived
(calculated) from sensor data.

11.4 Digital Inputs and Outputs

(8 pole female connection "In/Out")

- Output voltage (+U

B out

):

+U

B in

– 0.5 V

-

Current supply (incl. outputs):

max. 500 mA at 50 °C

- Inputs

Quantity:

4

Input voltage:

0 ... 35.0 V DC

Trigger threshold:

Low: < 0.8 V; High: > 5.0 V

Input current consumption:

approx. 4 mA

- Outputs

Quantity:

2

Switching capacity (per output):

+

U

B Out

/ 0.2 A

11.5 Interfaces

HSI bus

Mobile radio network:

GSM 850 / 900 (2W (EGSM))

GSM 1800 / 1900 (1W (EGSM))

Antenna

50

Ω FME plug

SIM:

3V SIM card

Advertising