Operation of the UL5000

41

iina

74

e 0

5

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

5

Operation of the UL5000

The UL5000 is switched on by pushing the mains switch (See Chapter

3.2.1

Startup

and Measure

). After about 2.5 min the run-up procedure is finished; the unit is in

Stand-by-mode and ready to measure.

Please connect the part to be tested to the inlet port and press START. The UL5000
starts to evacuate the part. The evacuation time depends on the volume of the test
part. During evacuation the screen shows the inlet pressure online.

Once the pressure of 15 mbar (11 Torr or 1500 Pa) is reached the unit switches to
measurement mode. The corresponding leak rate is displayed. For further explana-
tions of the screen please refer to

Fig. 5-1

.

The displayed leak rate corresponds to the helium background concentration in the
part under test. Since the UL5000 continues to pump down the part this background
leak rate will further reduce. As soon as the leak rate is low enough in respect to your
requirements you may start spraying Helium to search for possible leaks.

When you are finished please press STOP and hold the button a few seconds to vent
the part under test.

5.1

Display

The display is used to either show leak rate signals or program specific set-ups and
get information by means of the software menu (See Chapter

6

Description of the

Menu

). In addition messages and maintenance instructions are displayed on the

screen (See Chapter

8

Error And Warning Messages

).

5.2

The Screen in Run-Up Mode

In run-up mode the display shows:

Speed of the turbo molecular pump

Foreline pressure

State of emission

Active filament

A bar graph which shows the run-up progress

Notice: If the display is too bright or too dark, please refer to Chapter

6.2.4

.

5.3

Display in stand-by mode

In stand by mode the states are shown in the lower edge of the display (See Chapter

5.4.3

). Furthermore calibration (See Chapter

7

) can also be started in stand by mode

and purging, too (See Chapter

5.3.1

).