INFICON UL5000 Dry Helium Leak Detector User Manual

Page 79

Advertising
background image

Calibration

79

iina

74

e 0

7

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

After

External Calibration

(Soft Key no. 8) has been chosen the following messages

are displayed and the described actions are required:

Make sure that the test leak is connected and opened.

Check the leak rate printed on the test leak and compare it with the leak rate at
the display. If the leak rates are not identical press Edit leak rate (Soft Key no.
4) and correct the value.

If the leak rates are okay press START (Soft Key no. 8).

Fig. 7-1:

External Calibration

, Step 1

No action required.

Fig. 7-2:

External Calibration

, Step 2

The bargraph display shows a signal which must not vary much. If so please
press OK (Soft Key no. 8).

Fig. 7-3:

External Calibration

, Step 3