6.4.2

Trigger Level 2

Main Menu

>

Trigger & Alarms

>

Trigger Level 2

The value of the second trigger level can be set.

See Chapter

4.2.2.7

Numerical

Entries

for the description of the entry.

If Trigger 2 is exceeded the corresponding relay will switch. This is also indicated at
the display.

6.4.3

Volume

Main Menu

>

Trigger & Alarms

>

Volume

The minimum loudness and the regular volume of the loudspeaker can be adjusted.

The minimum loudness is the minimum speaker volume that cannot be exceeded to
even lower values. Thus it is avoided that the actual volume is accidentally reduced
to a value that is below the noise level of the environment.

The actual volume can be adjusted between 15 (maximum) and the value defined as
minimum loudness.

Softkey 2:

Arrow down

Decrease the minimum loudness. The minimum value is 0.

Softkey 3:

Arrow down

Decrease the actual volume. The minimum value is the value defined under
minimum volume.

Softkey 4:

Beep off / Beep on

Softkey 5:

Help for volume

Help

Softkey 6:

Arrow up

Increase the minimum volume. The maximum value is 15.

Softkey 7:

Arrow up

Increase the regular volume. The maximum value is 15.