6.3

Mode

Main Menu

>

Mode

The mode menu

Fig. 6-4

enables the submenu to select the different working modes.

Fig. 6-4: The

Mode

Menu

Key No.

Name

Description

1

Cancel

Return to the main menu without any changes
of the present settings.

7

Vacuum

The normal vacuum mode will be used.
Vacuum mode with the possibility to use
HYDRO•S.
HYDRO•S improves your time to test. Water
vapor desorption in a test object generates an
unwanted background signal. HYDRO•S
separates and eliminates this signal from the
helium signal, so testing can begin sooner.
After changing to this mode, a new calibration
necessary. Please refer to Chapter

6.6.1.4

,

HYDRO•S automatic ON.

5

Not used in this menu.

3

Sniff

The sniff mode will be used. Please refer to
Chapter

4.3.2

Sniffer Mode

.

8

OK

Save the settings and return to the previous
menu.