Trigger & alarms, Trigger level 1, To chapter – INFICON UL5000 Dry Helium Leak Detector User Manual

Page 52: See 6.4), 4 trigger & alarms, 1 trigger level 1

Advertising
background image

52

Description of the Menu

iina

74

e 0

6

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

6.4

Trigger & Alarms

Main Menu

>

Trigger & Alarms

The trigger levels, the volume of the loudspeaker and the units of leak rates and
pressures can be set in this menu

Fig. 6-5

.

6.4.1

Trigger Level 1

Main Menu

>

Trigger & Alarms

>

Trigger Level 1

The value of the first trigger level can be set.

See Chapter

4.2.2.7

Numerical Entries

for the description of the entry.

Trigger 1 and Trigger 2 are programmable leak rate levels. If these levels are
exceeded some outputs of the UL5000 will react, especially if Trigger 1 is exceeded.

Display

In the status line of the display the signs for Trigger 1 and Trigger 2 are displayed
inverted if the leak rate exceeds (becomes higher than) the programmed value.

Relay Output

The relays on the I/O interface

Fig. 2-6

/2 will switch.

See Chapter

2.3.2.2

Digital Out

for further details.

Alarm/Loudspeaker

Trigger level 1 defines at which level the various alarm types react (

See Chapter

6.4.6

Audio alarm type

)

Fig. 6-5: The

Trigger & Alarms

Menu

Key No.

Name

Description

1

Back

Return to the main menu.

2

Trigger Level 1

Definition of Trigger level 1. Please refer to Chapter

6.4.1

3

Trigger Level 2

Definition of Trigger level 2. Please refer to Chapter

6.4.2

4

Volume

Please refer to Chapter

6.4.3

5

Units

Selection of leak rate and pressure units. Please refer to
Chapter

6.4.4

6

Not used in this menu.

7

Alarm delay

Please refer to Chapter

6.4.5

8

Audio alarm type

Choice of different alarm types. Please refer to Chapter

6.4.6