4000 hours maintenance

The 4000 hours maintenance should be performed by an INFICON service tech-
nician or another authorised person at least yearly.

Independently of 4000 working hours the lubricant reservoir of the Turbomolecular
pump and the oil in the Fore pump should be replaced at least yearly.

Notice: The internal Helium standard leak certificate is valid for one year after

delivery. The annual refurbishing of the internal Helium standard leak is
recommended and an other certificate will be delivered. The internal
Helium standard leak can be refurbished at INFICON GmbH in Cologne
exclusively.

The maintenance work will take approximately 3 hours.

Work to be performed

Required materials

P/N

Replace the Lubricant Reservoir of the
Turbomolecular Pump TMH 071
(every 2 years)

Lubricant Reservoir for TMH 071

200000577

Check and/or replace Filters

Spare Filter Cell for the fans

Spare Filter for venting and purge line

Silencer for exhaust (only UL1000 Fab)

Filter cell for oil filter.

200000685

200000683

20099183

200000694

Functional check and adjustement