6.6.1.4

HYDRO•S

Main Menu

>

Settings

>

Vacuum settings

>

HYDRO•S

Notice: This setting is only valid, if mass mode is set to „

4

He“. See Chapter

6.3

.

You can always switch HYDRO•S on and off manually, even if automatic is enabled.

Notice: Switching HYDRO•S manually, deactivates the automatic until the next

measuring cycle (STOP/START).

6.6.1.5

Leak rate internal test leak

Main Menu

>

Settings

>

Vacuum settings

>

Leak rate internal test leak

The value of the internal test leak can be set. See Chapter

4.2.2.7

Numerical Entries

for the description of the entry.

Softkey 7:

enable

If HYDRO•S automatic is enabled, the UL5000 switches HYDRO•S on and
off at the following conditions:

HYDRO•S is switched on, if

ZERO is off and

p1<0,3mbar and

Leak rate is between 2×10

-10

mbar l/s and 3×10

-10

mbar l/s and

HYDRO•S was not switched on manually

The lower detecting limit for HYDRO•S is 1×10

-10

mbar l/s

Softkey 3:

disable

HYDRO•S is disabled. Leak rate higher than 3×10

-10

mbar l/s

If HYDRO•S is disabled, it can not be switched on/off manually anytime.

Softkey 5:

Automatic OFF

HYDRO•S can be switched on/off manually by Softkey 4 in display mode
(See Chapter

5.4.6

)

Warning

Normally there is no reason to edit the leak rate of the internal test leak besides after
a change of the internal test leak. A wrong leak rate of the internal test leak will lead
to wrong leak rate readings!