Vent delay, Vacuum ranges, See chapter – INFICON UL5000 Dry Helium Leak Detector User Manual

Page 59: To select

Advertising
background image

Description of the Menu

59

iina

74

e 0

6

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

6.6.1.2

Vent delay

Main Menu

>

Settings

>

Vacuum settings

>

Vent delay

Through this menu item it is possible to define the delay time until the inlet port is
vented when operating the STOP button. When the STOP button is pressed for a
period of time which is shorter than the delay time specified here, the UL5000 will just
change to Stand-by mode.

When the STOP button is pressed for a period of time which is longer than the delay
time specified here, the UL5000 will vent the inlet port.

6.6.1.3

Vacuum ranges

Main Menu

>

Settings

>

Vacuum settings

>

Vacuum ranges

With this menu you can adjust different modes concerning the activity of leak
detection. This setting is only active in mode vacuum (see Chapter

6.3

).

Softkey 2:

Immediately

The inlet port will be vented immediately after pressing the STOP button.

Softkey 3:

After 1 second

The inlet port will be vented with a time delay of 1 second.

Softkey 4:

After 1.5 seconds

The inlet port will be vented with a time delay of 1.5 seconds.

Softkey 6:

after 2 seconds

The inlet port will be vented with a time delay of 2 seconds.

Softkey 7:

No vent

The inlet port cannot be vented with the STOP button.

Softkey No. 2: ULTRA ONLY

In this mode the UL5000 remains in the area ULTRA after running under
0,4 mbar at the inlet flange (see Chapter

4.3.1

). When showing the pressure at

the inlet flange > 0,4 mbar the UL5000 switches immediately into evacuation
mode.

Softkey No. 3: FINE only

In this mode the UL5000 remains after falling below 2 mbar at the inlet flange in
FINE mode. Valve V1a will be closed. When the pressure at the inlet flange is
increasing > 1 mbar the UL5000 switches immediately into evacuation mode.
The lower detection limit of FINE ONLY is 1×10

-10

mbar l/s.

The advantage of FINE ONLY is that while this mode is running no valve will
switch.

Softkey No. 5: ?

Help

Softkey No. 7: all areas (default settings)

This is the default setting. The activity runs as explained in Chapter

4.3.1

.