6.6.6.1

Save parameter set

Main Menu

>

Settings

>

Parameter save / load

> Save

Save the current parameter settings.

6.6.6.2

Load parameter set

Main Menu

>

Settings

>

Parameter save / load

>

Save parameter set

The settings of the

selected

saved parameter set will be displayed and can be

reloaded.

6.6.7

Monitoring functions

Main Menu

>

Settings

>

Monitoring functions

Calibration request

Main Menu

>

Settings

>

Monitoring functions

> Calibration request

It can be selected whether the operator is reminded of the fact that a calibration may
has become necessary or not. The default value is off.

If the calibration request is switched on, a corresponding message will appear when
30 minutes have elapsed after power on or if the temperature of the UL5000 has
changed by more than 5 °C (9 °F) since the last calibration.

Particle protection

Main Menu

>

Settings

>

Monitoring functions

> Particle protection

This mode can be switched on and off.

If switched on the UL5000 will not start pumping down before the inlet ha

s not

dropped below

1

mbar. I. e. it is assumed that the part under test is pumped by

another pump in parallel.

Purpose: When the leak detector does not pump at high pressures no gas stream

,

possibly carrying particles gets into the leak detector.

Softkey 4:

Edit a file name

Rename the parameter set.

Softkey 6:

Arrow up

Upward to the previous screen.

Softkey 7:

Arrow down

Downward to the next screen.

Softkey 3:

Off

The calibration request will be switched off.

Softkey 7:

ON

The calibration request will be switched on.