8000 hours maintenance – INFICON UL5000 Dry Helium Leak Detector User Manual

Page 93

Advertising
background image

Maintenance works

93

iina

74

e 0

9

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

9.5.3

8000 hours maintenance

The 8000 hours maintenance should be performed by an INFICON service tech-
nician or an other authorized person.

The scroll module of the scroll pump IWATA should be replaced every 8000 working
hours, at least every 2 years.

After 8000 working hours, the „Tip Seal“ of the scroll module of the Varian pump
should be replaced by an INFICON service technician. If the „Tip Seal“ has not been
replaced, then after 12000 working hours the scroll module must be exchanged.

The maintenance work will take approximately 6.0 hours.

Work to be performed

Required materials

P/N

Exchange the scroll module (IWATA ISP-
500B)

Scroll module for IWATA ISP-500B

200000217R

Replace the Tip Seal (Varian TS 620)

Tip Seal

2000001671

Restore the Lubricant Reservoir TMH 071 •

Lubricant Reservoir for TMH 071

200000577

Replace the seals for valves

Set of seals for valves

200000594

Check and/or replace Filters

Spare Filter Cell for the fans

Spare Filter for venting and purge line

Silencer for exhaust

200000685

200000683

20099183

Functional check and adjustement

Advertising