Description of the maintenance work

Only trained specialist staff can performe more changes at the UL5000 than the
normally maintenance work.

9.6.1

Opening the UL5000

Required tool

Wedge (Accessories).

Separate the UL5000 from other vacuum components at the inlet port.

Remove the side covers by using the wedge (

Fig. 9-1

). Push down the wedge to

release the side covers.

The location for the wedge is marked with two dot marks at the top side of the side
covers (See

Fig. 9-1

/2).

Open the both side covers in the same way.

STOP

Danger

The protective conductor screw at the chassis plate should not be loosened. The
operator is not protected against electric shock by working without a protective
conductor line.

STOP

Danger

Disconnect the power cord from the UL5000 before opening the side cover.