INFICON UL5000 Dry Helium Leak Detector User Manual

Page 8

Advertising
background image

8

General Information

iina

74

e 0

1

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

Auto zero

Determination and automatic adaptation of the helium background.

Through this function, the internal zero level of the instrument is determined which is
then subtracted from the current leak rate signal. This function is run during the cali-
bration process or when operating the start pushbutton, provided the UL5000 has
been running previously for at least 20 seconds in the standby or vent mode. If the
helium background previously suppressed should drop so that for the duration of the
zero time only the display limit will be displayed, the zero level is adapted automati-
cally.

Menu

The menu allows the user to program the UL5000 according to his requirements. The
menu has a tree architecture.

Default

Status of the UL5000 when supplied by the factory.

GROSS

GROSS is a measurement mode which allows high inlet pressure (1 to 15 mbar).
The smallest detectable leak rate is 1×10

-6

mbar l/s.

FINE

FINE is the medium measurement mode with inlet pressure between 2 and 0,4 mbar.
Detection limit is 1×10

-10

mbar l/s.

ULTRA

ULTRA is the most sensitive measuring range with inlet pressures below 0,4 mbar.
The minimum detectable leak rate is < 5×10

-12

mbar l/s.

Foreline pressure

Pressure in the foreline between Turbo pump and scroll pump.

Minimum detectable leak rate

The smallest leak rate the UL5000 is able to detect (

 5E-12 mbar l/s).

Internal helium background

The existing helium partial pressure in the measurement system. The level of the
internal helium background is measured in the Stand-by mode and subtracted from
the measured signal.

Measure Measurement mode

The UL5000 measures the leak rate of the test sample.