Maintenance groups

The maintenance plan for the UL5000 can be subdivided in 4 maintenance groups.

1500 hours maintenance

4000 hours maintenance

8000 hours maintenance

16000 hours maintenance

9.5.1

1500 hours maintenance

The 1500 hours maintenance can be performed by an operator or a maintenance
person.

The filter cell in front of the fans should be checked and replaced if dirty. By operating
under worth conditions, the maintenance rates can be appropriatly reduced.

Replace the silencer at the exhaust of the leak detector.

Notice: An obstructed silencer can lead to damaging the scroll pump

Work to be performed

Required materials

P/N

Check and/or replace filters

Replace silencer

Spare filter cell for the fans

Silencer for the exhaust

200000685

20099183