Audio alarm type

Main Menu

>

Trigger & Alarms

>

Audio alarm type

The trigger of the audio alarm can be switched on or off.

6.4.6.1

Pinpoint

The tone of the acoustical signal changes its frequency only in a LR-window

Fig. 6-

6

which ranges from one decade below the Trigger level 1 up to one decade above

the Trigger level 1. Below the window the tone is constantly

low

, above the window

it is constantly high.

Example: The Trigger level 1 is 4×10

-7

mbar l/s. So the window where the tone

changes reaches from 4×10

-8

mbar l/s up to 4×10

-6

mbar l/s.

6.4.6.2

Leak rate prop.

The frequency of the accoustic output is proportional to the reading on the bargraph
display. The frequency ranges from 200 Hz to 1300 Hz. See Chapter

6.2.1

Scale

linear/logarithmic

for the definition of the number of decades.

Softkey 2:

Pinpoint

See Chapter

6.4.6.1

Softkey 3:

Leak rate prop.

The sound will be proportional to the leak rate signal.

See Chapter

6.4.6.2

Softkey 6:

Setpoint

See Chapter

6.4.6.3

Softkey 7:

Trigger alarm

An alarm sounds when the trigger 1 is exceeded.

See Chapter

6.4.6.4