6.6.4.4

Scaling Recorder Output

Main Menu

>

Settings

>

Interfaces

>

Scaling Recorder Output

Here the scaling of the recorder output can be adjusted. This adjustment is only valid
for the setting “LR lin” or ”LR log” (refer to Chapter

6.6.4.2

Recorder output

).

Example:
Chart recorder output: “LRlog”
Upper limit value is adjusted to 10

-5

(= 10V)

Scaled to 5 V /decade
Lower limit value consequently is 10

-7

(= 0 V)

Softkey 5:

?

Help

Softkey 6:

UL2xx Leak Ware

Gives the chance to control and read measurement values when connecting to a
computer.

Notice: The calibration function of the Leak Ware is not appropriate to operate with

the UL5000.
Please execute the function „STORE DATE“ in the operating mode „Single
Part Measurement“ for starting the record of the measured values.

Softkey 7:

ASCII

Gives the chance to use the UL5000 via a RS232 terminal.

Softkey 2:

Arrow up

Adjust decade of the upper limit value

Softkey No. 3: Arrow down

Scaling of the previously adjusted value in steps of 0.5, 1, 2, 2.5, 5, 10 Volt/decade. The
complete voltage range is 10 V. (Only for signal “LRlog”)

Softkey No. 6: Arrow up

Adjust decade of the upper limit value

Softkey No. 7: Arrow down

Scaling of the previously adjusted value in steps of 0.5, 1, 2, 2.5, 5, 10 Volt/decade. The
complete voltage range is 10 V. (Only for signal “LRlog”)