1 outputs tab – INFICON SQM-242 Thin Film Deposition Controller Card Software Operating Manual User Manual
Page 51
3 - 21
IP
N 07
4-
55
1-
P1
A
SQS-242 Operating Manual
3.5.4.1 Outputs Tab
Figure 3-18 Outputs tab
Name: Assigns a name to each displayed output. For clear display, keep the name
to less than 8 characters.
Physical Output: Up to six SQM-242 cards (physical outputs 1 to 12) and a single
SAM-242 card (physical outputs 13 and 14) may be installed in a computer.
However, the SQS-242 software can display and control a maximum of 6 outputs
simultaneously. Use this dropdown to assign a “physical” output to a “display”
output.
FS Out: The input voltage required by the deposition source power supply to
produce 100% output power. Positive or negative full scale values are possible.
Color: Selects the color used to graph and display output data.
Test Output: Useful for testing output wiring and Full Scale voltage settings. Select
an output, then click Full Scale to set the SQM-242 card output to its Full Scale
voltage. Click Zero to return the selected output to 0 volts.