3 - 21

IP

N 07

4-

55

1-

P1

A

SQS-242 Operating Manual

3.5.4.1 Outputs Tab

Figure 3-18 Outputs tab

Name: Assigns a name to each displayed output. For clear display, keep the name
to less than 8 characters.

Physical Output: Up to six SQM-242 cards (physical outputs 1 to 12) and a single
SAM-242 card (physical outputs 13 and 14) may be installed in a computer.
However, the SQS-242 software can display and control a maximum of 6 outputs
simultaneously. Use this dropdown to assign a “physical” output to a “display”
output.

FS Out: The input voltage required by the deposition source power supply to
produce 100% output power. Positive or negative full scale values are possible.

Color: Selects the color used to graph and display output data.

Test Output: Useful for testing output wiring and Full Scale voltage settings. Select
an output, then click Full Scale to set the SQM-242 card output to its Full Scale
voltage. Click Zero to return the selected output to 0 volts.