6 update utility, 7 query measurement, 8 query register – INFICON SQM-242 Thin Film Deposition Controller Card Software Operating Manual User Manual
Page 82: 9 update register
5 - 8
IP
N 07
4-
55
1-
P1
A
SQS-242 Operating Manual
5.8.6 Update Utility
@QU;<param1>;<Index>
02
SQM242 Mode (0/1)
03
SQM242 Period
04
SQM242 Filter
06
Front Panel Enabled (0/1)
07
Application Visible (0/1)
08
Full Scale Output (Index=1 to 6)
14
Application On Top(0/1)
5.8.7 Query Measurement
@QM;<param1>
1n
Output n Power (n=1 to 6)
2n
Output n Rate
3n
Output n Thickness
4n
Output n Deviation
5n
Sensor n Rate (n=1 to 8)
6n
Sensor n Thickness
7n
Sensor n Frequency
8n
Sensor n % Life
9n
Analog Input n Voltage
5.8.8 Query Register
@QR;<param1> (returns register value in HEX)
0
Layer/Phase Register
1
Sensor/Output Register
2
Analog/Output Register
20
Source Index Register
21
Source Index Done Register
22
Relay Register
24
Input Register
25
Layer Index Register
26
Layer Index Done Register
5.8.9 Update Register
@UR;<param1>;<value> (value in HEX)
0
Layer/Phase Register
1
Sensor/Output Register
2
Analog/Output Register
20
Source Index Register
21
Source Index Done Register
22
Relay Register
24
Input Register
25
Layer Index Register
26
Layer Index Done Register
XXX
Any Register