Overview

1/12 Oct + Tol:

The limit button lights up green for results

within tolerance and red for out of tolerance results.

5

Pause

Pauses the current measurement. Press the pause button

or the start/stop button

to continue the measure-

ment. The data logging is continued in the function SLMe-
ter/RTA
during the pause period and listed in the log file.

6

Start/Stop

Starts and stops a measurement.

7

Enter

Confirms a selection.

8

Rotary Wheel

Selects the required measurement function or the individu-
al measurement parameter.

9

ESC

Terminates an entry, returns to the top menu level or closes
an open window.

Hold the speaker button

down and adjust the level

of the speaker and headphone outputs with the rotary
wheel . Digital gain control prevents gain increases be-
yond a level that will cause signal clipping. The maximum
gain setting depends on the connected input signal.
The headphone output provides a linear output signal

using the SLMeter function (for input levels of approx.
-30 dBu and higher, which equals at M2230: 85 dBA,
M2211: 92 dBA utilizing a pink noise test signal).

3

Power & Backlight

The power button

switches the instrument on. The XL2 is

immediately ready for operation. Holding down the power
button

for one second switches the XL2 off. Additionally,

a brief press of the power button

toggles the backlight

during operation.

4

Limit

SLMeter:

The limit button lights up green, yellow or red

according to the settings in the Limit page. Press the
limit button

to access the Limit page. For more details

visit the chapter Sound Level Meter: Limits.
FFT + Tol:

The limit button lights up green for results

within tolerance and red for out-of-tolerance results.
Polarity:

The limit button lights up green at positive polar-

ity and red at negative polarity.