132

Audio Analyzer

1

Filter

Z-Weighting

Frequency Z-weighting with flat frequen-
cy response from 20 Hz to 22 kHz. De-
fault measurement setting.

A-Weighting Frequency

A-weighting

according

IEC 61672 for low noise levels.

C-Weighting

Frequency

C-weighting

according

IEC 61672 for high noise levels.

HP 100Hz

Highpass 100 Hz, -200 dB/dec.,
maximally flat magnitude filter,
e.g. -60 dB @ 50 Hz

HP 400Hz

Highpass 400 Hz, -120 dB/dec., maxi-
mally flat magnitude filter, attenuates
50/60Hz components of test signal

HP 19k

Highpass 19 kHz; e.g., to measure a
20 kHz pilot tone level of a public address
system without disturbing the public.

22.4 - 22.4k

Bandpass filter 22.4 Hz - 22.4 kHz (IEC
468-4)

2

Level RMS

Measures the absolute level of the input signal. The unit’s
dBu, dBV, V and dBSPL are selectable.

3

THD+N

Measures the total harmonic distortion and noise of the
input signal. The distortion measurement, expressed in dB,
linear or in %, is conducted within the bandwidth 10 Hz -
20 kHz.

4

Frequency in Hz

The XL2 automatically extracts and measures the frequen-
cy of the fundamental signal. The XL2 frequency counter
technology even reads the frequency correctly for heavily
distorted signals.

5

Balance Indicator

Indication of the audio signal balance between pin 2 and 3
on the XLR input for input signals > -34 dBu.

The input signal is balanced.

The signal is unbalanced. The level at pin 2 is
higher than at pin 3.
The signal is unbalanced. The level at pin 3 is
higher than at pin 2.