Reverberation Time RT60

Start the Measurement

Press start

. The status indication switches to ARMED.

Enable the test signal; e.g., switch on the gated pink noise

signal or fire a starter pistol.
The black bars have to exceed the grey bars completely.

The RT60 calculation cycle is complete when

the sound pressure level falls below the bot-
tom of the grey level markers in every band.
Confirmation marks

, indicating success-

ful measurements, are displayed above each
band.

Continue the Measurement

Using pink noise:
The gated pink noise test signal continues with the preset on/
off time. The XL2 automatically triggers at each cycle. Com-
plete a minimum of three test cycles to accurately calculate
the uncertainty factor.

Using an impulse source:
Use further starter pistol shots to repeat the measurements.

Press pause

to pause the measurement.

Prepare the next impulse sound source; e.g., starter pistol.

Press start

to continue the reverberation time measure-

ment.
Enable the test signal; e.g., fire the starter pistol.

An automated algorithm calculates the averaged reverb time of
all measurement cycles.