Capture + tolerances – NTi Audio XL2 Sound Level Meter User Manual

Page 155

Advertising
background image

155

Capture + Tolerances

#Hide
UnusedBands

This parameter defines which tolerance fre-
quency range is shown. (not mandatory)

True

Measurement results of fre-
quency bands without any
specified tolerance are not dis-
played.

False

The measurement result of all
frequency bands are shown.

#nAllowed
Violations

Allows a PASSED result with a maximum
number of n failed frequency bands (not man-
datory)

#FreqScale
Spacing

Defines the frequency scale spacing in the X-
axis for the tolerance data. (not mandatory)

lin

Tolerance data displayed with
linear frequency spacing, e.g.
used in the FFT function. (de-
fault setting)

log

Tolerance data displayed with
logarithmic frequency spacing
e.g. used in the 1/12 Oct + Tol
function.

#AutoStart

Enables auto start function. (not mandatory)

True

Auto start is active. The run indi-
cation displays A. The measure-
ment starts automatically after
the trigger level is reached.
Alternatively the measurement
may start manually by pressing
start

.

False

No auto start function. (default
setting)

Advertising