Specifications – NTi Audio XL2 Sound Level Meter User Manual

Page 210

Advertising
background image

210

Specifications

Functions of

Spectral

Limits

Option

(optional)

SLMeter/RTA function

True peak level in 1/1 and 1/3 octave resolution

»

FFT function

High-resolution Zoom-FFT with selectable fre-

»

quency ranges and resolution up to 0.4 Hz in the

range of 5 Hz to 20 kHz

1/12 octave function

High resolution RTA function “1/12 Oct + Tol”

»

Selectable 1/1, 1/3, 1/6 and 1/12 octave spectral

»

resolution

Frequency band listening at rear speaker

»

FFT and 1/12 octave function

Capturing of multiple readings into the internal

»

memory

Comparing measurement results against cap-

»

tures with relative or absolute curve display

Comprehensive tolerance handling with toler-

»

ance masks based on captures for passed/failed

measurements

Export and import of tolerance and capture files

»

Noise Curves

Noise Rating NR in accordance with ISO 1996

»

Noise Criteria NC

»

in accordance with ANSI S12.2-2008 and -1995

Room Noise Criteria RNC

»

in accordance with ANSI S12.2-2008

Room Criteria RC

»

in accordance with ANSI S12.2-1995

Preferred Noise Criteria

»

in accordance with ASA 1971

Acoustic Analyzer
FFT Analysis

Real-time FFT with actual level, L

eq, Lmin, Lmax

Level resolution: 0.1 dB

Ranges: 7 Hz - 215 Hz, 58 Hz - 1.72 kHz, 484

Hz - 20.5 kHz with 142 frequency bins shown on

display

Measurement Units: Volt, dBu, dBV and dBSPL

Optional: High-resolution Zoom-FFT with select-

able frequency ranges and resolution up to 0.4 Hz

in the range of 5 Hz to 20 kHz

Optional: Capture and tolerance function with mul-

tiple readings for comparative measurements and

passed/failed analysis

Reverbera-

tion

Time RT60

1/1 octave bands results from 63 Hz - 8 kHz,

based on T20

Optional: 1/3 octave bands results from

50 Hz - 10 kHz, based on T20

Range: 10 ms - 14 seconds

Measurement in accordance with ISO 3382 using

Schroeder-method

Test signal: Impulse source or gated pink noise

generated by the MR-PRO, MR2 or the included

NTi Audio Test CD

Delay Time

Propagation delay between electrical reference

signal and acoustic signal using the internal micro-

phone

Range: 0 ms - 1 second (0 m - 344 m)

Resolution: 0.1 ms

Test signal: NTi Audio delay test signal generated

by the MR-PRO, MR2 or the included NTi Audio

Test CD