Capture + Tolerances

Passed Result

The tolerance mode is active for passed/failed measure-
ments.

The passed/failed condition is further visualized by the bicolor
“limit” button and forwarded to the I/O interface of the instru-
ment in order to drive an external alarm device such the acces-
sory SPL Stack Light.

2

3

1

1

PASSED / Failed Result

PASSED

The actual measurement result is within the
tolerance band.

The actual measurement result is outside of
the tolerance band. Exceptions are visual-
ized in any out-of-tolerance frequency band.

2

Tolerance Bandwidth

Adjusts the tolerance bandwidth based on the captured ref-
erence spectrum in dB.

3

Level Offset

Provides a level offset in dB to move the tolerance band
up/down.