STIPA

STIPA with Ambient Noise Correction

Measuring the speech intelligibility index under realistic en-
vironmental conditions is often not feasible; e.g., playing the
test signal in a railway station at emergency levels during peak
hours will irritate passengers. Additionally, at rush-hour the
characteristics of ambient noise might be highly impulsive,
while a pre-requisite for accurate STIPA measurements is a
negligible impulsivity in the ambient noise. Under such circum-
stances the STIPA measurement should be shifted to a more
suitable time of the day; e.g. night time.

Measurement Sequence

First measure the ambient noise

•

Secondly measure the speech intelligibility STIPA

•

This sequence simplifies the STIPA measure-
ment as follows: The XL2 immediately displays
the STIPA result with ambient noise correction.
This result provides a guideline if repeated mea-
surements at the same location with averaging
is required. For details see the chapter STIPA
Measurement Hints.

Enable Ambient Noise Correction

Select the correction page

•

Cor.

Turn the rotary wheel

•

to select the parameter Ambient

Noise Correction and press enter

.

The correction enables and the XL2 displays this screen: