STIPA

STIPA with Ambient Noise Correction

Measuring the speech intelligibility index under realistic en-
vironmental conditions is often not feasible; e.g., playing the
test signal in a railway station at emergency levels during peak
hours will irritate passengers. Additionally, at rush-hour the
characteristics of ambient noise might be highly impulsive,
while a pre-requisite for accurate STIPA measurements is a
negligible impulsivity in the ambient noise. Under such circum-
stances the STIPA measurement should be shifted to a more
suitable time of the day; e.g. night time.

Measurement Sequence

First measure the ambient noise

Secondly measure the speech intelligibility STIPA

This sequence simplifies the STIPA measure-
ment as follows: The XL2 immediately displays
the STIPA result with ambient noise correction.
This result provides a guideline if repeated mea-
surements at the same location with averaging
is required. For details see the chapter STIPA
Measurement Hints.

Enable Ambient Noise Correction

Select the correction page

Cor.

Turn the rotary wheel

to select the parameter Ambient

Noise Correction and press enter

.

The correction enables and the XL2 displays this screen: