Microphones

Microphone plugs directly into the XL2

The XL2 automatically reads the electronic data sheet of the
connected microphone as follows:

Connect the measurement microphone to the XL2.

Switch on the XL2.

The XL2 reads the electronic data sheet of the connected

microphone during a brief initialization process prior to the
first measurement.

Microphone Connection via the ASD Cable

The NTi Audio measurement microphones can be connected
to the XL2 Analyzer using an ASD Cable for measurements at
remote locations or for reducing acoustic reflections. The elec-
tronic data sheet is transmitted via the XLR connector’s hous-
ing. Do not touch this during the brief initialization period to
ensure the complete data sheet is recognized by the XL2. The
automated sensor detection does not disturb any measure-
ments. You may join up to 4x ASD Cables together in series.
The ASD technology supports accurate data communication up
to a combined cable length of 20 meters (= 65 feet).

Microphone Connection via a professional Audio Cable

For distances longer than 20 meter (= 65 feet) use a high qual-
ity, low capacitance standard professional audio cable. The mi-
crophone sensitivity has to be entered manually into the XL2
Analyzer.

Alternatively connect the microphone first directly to the Ana-
lyzer. The XL2 reads the sensitivity and remembers this value.
Afterwards connect the audio cable.

Use the microphone for the intended

purpose only.
Protect the microphone from contamination

by always using the supplied windscreen.
Never use the microphone in a damp or wet

environment.
Do not jar or drop the microphone.

Do not remove the microphone protective grid.

Do not touch the microphone membrane.

Remove the black dust cap of the 1/2”

measurement microphones prior to use.