Capture + Tolerances

Advanced Start Functions

You may start the passed/failed measurement by one of the
following possibilities:

Press the XL2 start button

•

.

Activate the automated level trigger in the tolerance file.

•

Activate the digital input 1 via the external I/O interface; e.g.

•

a foot switch or a PLC control.

Digital I/O Interface

Result Passed:

Output 1

Result Failed:

Output 3

Start Measurement:

Input 1

Tolerance Management

Tolerance curves can either be imported from txt-files or direct-
ly derived from captured measurements. The XL2 calculates
tolerance bands based on

Single captured results

•

Manually generated txt-files on the PC

•

Mean average of multiple captures

•

Min/Max curves of multiple captures

•

Create Tolerance Files Manually on PC

Capture a spectrum on the XL2 Analyzer.

•

Enter the tolerance mode and generate a tolerance band for

•

the capture.
Select the Capture & Start Tolerance Mode symbol.

•

Confirm with enter

•

.

Select

•

Save tol. to SD Card to export the capture tolerance

file. This file includes the reference capture and the tolerance
data. The XL2 generates a txt-file in the folder “Tolerances”.
Open the capture tolerance file with a text editor or MS Ex-

•

cel.
Edit the tolerance data according your requirements. You

•

may delete any unused frequencies. All remaining frequency
data has to remain the same. Any frequency changes will be
discarded by the XL2.