Fft analysis + tolerance – NTi Audio XL2 Sound Level Meter User Manual

Page 66

Advertising
background image

66

FFT Analysis + Tolerance

3

Y-Scale setting

Select the Y-Axis with the rotary wheel

and confirm

with enter .
Select the zoom factor between

20, 10, 5, 2.5 dB/div

and confirm with enter .
Scroll up and down with the rotary wheel

to select the

Y-axis range.
Confirm with enter

.

4

Measurement Result

Displays the actual and averaged measurement results.

5

Input Range

The XL2 Analyzer provides three input ranges to accom-
modate the wide range of input signals. The individual rang-
es are based on the microphone sensitivity setting in the
calibration menu of the XL2. For example at a sensitivity
S=20mV/Pa the input ranges are

Lower range: 10 - 110 dBSPL

Middle range: 30 - 130 dBSPL

Upper range: 50 - 150 dBSPL

Select the lowest possible input range according to the
maximum level expected during the measurement; e.g., if
the sound pressure will be always below 110 dBSPL, then
select the lowest input range 10 - 110 dBSPL.

Manage
captures

Allows to rename captures, clear re-
corded captures, save captures to SD-
Card for export from the XL2 or load
captures from SD-Card for import to
the XL2.

Start tolerance
mode

Starts the tolerance mode for passed/
failed measurements comparing the
actual measurement results against a
tolerance band.

2

Measurement Result

Actual level result of the indicated frequency band. The cur-
sor readout displays the center frequency and the level of
the band indicated by the arrow.

Upper parameter displayed as line.

Lower parameter displayed as bargraph.