Reverberation Time RT60

4

Measurement Status STAT

Displays the actual measurement status. Start and stop the
reverberation time measurement with start/stop

. The

following status information is displayed:

ARMED

Measurement is waiting for the test signal
to exceed at least one measurement trigger
marker, then the measurement is triggered
automatically.

NOISE

A sound stimulus signal above the trigger
marker is being received.

DECAY

Decay is being measured.

PAUSE

The measurement has been paused by press-
ing pause

.

STOP

No RT60 measurement is presently being per-
formed.

5

Average AVRG

Counts the measurement cycles. When using a gated
noise test signal, a minimum of 3 sequential cycles is rec-
ommended.

6

Actual Real-Time Spectrum

The black bars indicate the actual signal, including back-
ground noise. While setting up the reverberation time mea-
surement you should increase the test signal level until the
black bars exceed the grey bars completely and the status

4

displays NOISE.

7

Level Marker

The grey bars indicate the signal level required for effective
reverberation time measurements in each octave band.
The marker has a length of 35 dB, and can be set by

Selecting

SET

1

with the rotary wheel .

Press enter

to measure the environmental noise in

the silent room.

8

Band Status

Indicates a successful reverberation time measurement
above each octave band with a tick .

9

Measurement Resolution

Reverberation time measurement selectable in 1/3 oc-
tave or 1/1 octave band resolution. The optional Extended
Acoustic Pack enables 1/3 octave band resolution from
50 Hz to 10 kHz.