FFT Analysis + Tolerance

Save the FFT Measurement Result

Select the memory menu

with the rotary wheel and

confirm with enter .

The memory menu opens.

Press enter

to store the test results or select Save

Screenshot to store the screenshot.

The FFT measurement is completed.

Capture References and Create Tolerances

The Spectral Limits Option extends the function range of the
XL2 with trace capturing, relative curve display and compre-
hensive tolerance handling for the FFT Analysis and the high
resolution 1/12 Oct + Tol spectral analysis.

Features:

Captures multiple traces in the internal memory

Comparing measurement results against captured traces

with relative or absolute curve display
Comprehensive tolerance handling

Creating tolerance masks based on captured reference

curves for passed / failed measurements
Export and import of tolerance and capture files

True peak level in 1/1 and 1/3 octave resolution

High-resolution Zoom-FFT up to 0.4 Hz steps in the frequen-

cy range 5 Hz - 20 kHz

Read the detailed description in the separate chapter
Capture + Tolerance.