STIPA

1

Run Indication

This icon displays the run status of the STIPA measure-
ment.

2

Measurement Result

Single value speech transmission index result.

Repeatability of measurement result:

Since a dedicated noise test signal is used, the STIPA
result may deviate by maximum 0.03 STI (=Max-Min) at
the same measurement position.

3

Sound Level LAeq

Shows the time-averaged sound level of the 15 seconds
STIPA measurement cycle time.

4

Sound Level LAS

Actual sound pressure level.

5

Analog STIPA Bargraph

Bargraph display and interpretation of the speech intelligi-
bility measurement result

Exlnt

0.75 - 1.00 STI

Good

0.60 - 0.75 STI

Fair

0.45 - 0.60 STI

Poor

0.30 - 0.45 STI

Bad

0.00 - 0.30 STI

6

Qualification Scale

The STI value is shown as a letter of the qualification scale
below, which informs about the typical STI requirements
for dedicated applications.

Band

STI Range

Examples of typical uses

A+

> 0.76

recording studios

A

0.72 - 0.76

theatres, speech auditoria,

parliaments, courts

B

0.68 - 0.72

theatres, speech auditoria,

parliaments, courts

C

0.64 - 0.68

teleconference, theatres

D

0.60 - 0.64

class rooms, concert halls

E

0.56 - 0.60

concert halls, modern churches

F

0.52 - 0.56

PA in shopping malls,

public offices, cathedrals

G

0.48 - 0.52

PA in shopping malls,

public offices

H

0.44 - 0.48

PA in difficult acoustic

environments

I

0.40 - 0.44

PA in very difficult spaces

J

0.36 - 0.40

not suitable for PA systems

U

< 0.36

not suitable for PA systems