Noise curves – NTi Audio XL2 Sound Level Meter User Manual

Page 112

Advertising
background image

112

Noise Curves

12

SIL Measurement Result

The speech interference level (SIL) result is shown at noise
curve types NC-2008 and NC-1995

1

. It is calculated by

averaging the octave-band sound pressure levels at 500
Hz, 1000 Hz, 2000 Hz and 4000 Hz. If the measured spec-
trum in any octave band does not exceed any of the octave
bands of that NC(SIL) curve, the spectrum is designated
NC(SIL). If one or more octave-band levels exceed the
NC(SIL) curve, then the NC curve number may differ to the
SIL value, as the NC rating for that spectrum must be de-
termined using the tangency method.

13

Large Fluctuations

This reports any large fluctuations at low frequencies
for the Room Noise Criteria RNC in accordance with
ANSI S12.2-2008. If one or more displayed measurement
results exceed the limits, then large fluctuations exist. Con-
cluding the LEQ levels in the octave frequency bands be-
low 300 Hz are automatically “penalized” at the RNC mea-
surement.

14

Correction Values

The noise curve type RNC uses these correction values
based on the measured fluctuation results to determine
the RNC curve result.

Advertising