Calibration – NTi Audio XL2 Sound Level Meter User Manual

Page 137

Advertising
background image

137

Calibration

5

Level Type

LZF

Default setting.

LHP100F Suppressing any low frequency noise, for

example wind during outdoor calibrations.

6

Electronic Data Sheet of Measurement Microphone

Read out data from the connected microphone.

User Calibration - With External Precision Calibrator

User-sensitivity setting for measurement microphones with an
external calibrator.

Turn the rotary wheel

•

to select the parameter Calibration

Level

4

and press enter .

Set the calibration level according to the calibrator used.

•

Generate the specific reference signal with the calibrator

•

onto the microphone.
Turn the rotary wheel

•

to select the parameter RUN next

to User Calibration

3

and press enter .

The pop-up window

•

Calibration: Calibration running ...

shows up and changes to Calibration: Successfully fin-
ished!
for a successfully completed calibration.

The user calibration is completed.

User Calibration - Manual Sensitivity Setting

In case no external calibration device is available, set the sensi-
tivity of non-NTi Audio microphones manually according to the
product specifications as follows:

Select

•

Sensitivity

2

with the rotary wheel .

Press enter

•

and adjust the sensitivity with the rotary

wheel in 0.1 mV/Pa steps. The setting range 100 µV/Pa -
9.99 V/Pa is supported.
Confirm with enter

•

.

The manual user calibration is completed.

User Sensitivity

The XL2 writes the user-sensitivity setting onto
the electronic data sheet of the connected NTi
Audio measurement microphone or microphone
preamplifier. All subsequent measurements au-
tomatically utilize this user-sensitivity.