Sound level meter – NTi Audio XL2 Sound Level Meter User Manual

Page 60

Advertising
background image

60

Sound Level Meter

How to Measure the Correction Value

Play a pink noise signal at the typical sound pressure level of

the live event. (signal source Minirator, NTi Audio Test CD)
Select the correction page

.

Search the audience area for the location with the highest

sound pressure level and position the XL2 Analyzer.
Turn the rotary wheel

to select the parameter RUN next

to Audience and press enter .

The XL2 measures the sound level at the loudest position
for 5 seconds. The timer counts down to zero.

Wait until the measurement is completed.

Position the XL2 at the measurement position.

Turn the rotary wheel

to select the parameter RUN next

to Measure and press enter .

The XL2 measures the sound level at the measurement
position for 5 seconds. The timer counts down to zero.