How to Measure the Correction Value

Play a pink noise signal at the typical sound pressure level of

•

the live event. (signal source Minirator, NTi Audio Test CD)
Select the correction page

•

.

Search the audience area for the location with the highest

•

sound pressure level and position the XL2 Analyzer.
Turn the rotary wheel

•

to select the parameter RUN next

to Audience and press enter .

The XL2 measures the sound level at the loudest position
for 5 seconds. The timer counts down to zero.

Wait until the measurement is completed.

•

Position the XL2 at the measurement position.

•

Turn the rotary wheel

•

to select the parameter RUN next

to Measure and press enter .

The XL2 measures the sound level at the measurement
position for 5 seconds. The timer counts down to zero.