Metrohm 916 Ti-Touch User Manual
Page 181

■■■■■■■■■■■■■■■■■■■■■■
15 Templates
916 Ti-Touch
■■■■■■■■
165
[Edit pH value]
Enter the pH value for the selected temperature.
"Temperature table / Edit pH value" dialog
In this dialog, the pH values of the buffer are defined at various tempera-
tures. Enter the pH values for the temperature range in which you will
carry out your pH calibration and pH measurement. If you do not know
the pH values at individual temperatures, they will be calculated automati-
cally by means of linear interpolation.
Temperature in °C
Input range
0.0 - 95.0 °C (Increment: 5.0)
pH value
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the 32 Technical specifications chapter, page 429.
Input range
–20.000 - 20.000
Selection
off
Default value
off