28.7 Electrode test for pH electrodes (ELT)
■■■■■■■■■■■■■■■■■■■■■■
342
■■■■■■■■
916 Ti-Touch
■
"+": counterclockwise rotation
■
"–": clockwise rotation
Input range
–15 - 15
Default value
8
28.7.5
Further information
Table 10
Limit values of the three electrode types
Rating
Excellent electrode
Good electrode
Usable electrode
Electrode type "Standard"
Streaming potential
2.5 mV
3.0 mV
4.0 mV
Drift
2.0 mV/s
2.5 mV/s
3.0 mV/s
Min. slope
96.5 %
96.0 %
95.0 %
Max. slope
101.0 %
102.0 %
103.0 %
Response time
45 s
50 s
60 s
Lower limit Uoff
–15 mV
Upper limit Uoff
15 mV
Electrode type "Gel"
Streaming potential
3.0 mV
3.5 mV
4.5 mV
Drift
2.5 mV/s
3.0 mV/s
4.0 mV/s
Min. slope
96.5 %
96.0 %
95.0 %
Max. slope
101.0 %
102.0 %
103.0 %
Response time
60 s
75 s
90 s
Lower limit Uoff
–20 mV
Upper limit Uoff
20 mV
Electrode type "Non-aqueous"
Streaming potential
3.0 mV
4.5 mV
6.0 mV
Drift
5.0 mV/s
7.0 mV/s
9.0 mV/s
Min. slope
88.0 %
80.0 %
70.0 %
Max. slope
120.0 %
130.0 %
140.0 %
Response time
60 s
75 s
90 s
Lower limit Uoff
–50 mV
Upper limit Uoff
100 mV