8 stored buffer series for ph calibration – Metrohm 916 Ti-Touch User Manual

Page 431

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

31 Appendix

916 Ti-Touch

■■■■■■■■

415

31.8

Stored buffer series for pH calibration

The temperature-dependent pH values of several common pH buffers are
stored in the system for automatic buffer recognition during pH calibra-
tion. Apart from the Metrohm buffer solutions other reference and techni-
cal buffers are also included in the tables.

The following tables provide an overview of the stored pH(T) series.

pH values printed in bold are the values for the reference temperature of
the respective buffer set.

pH values printed in italics are interpolated or extrapolated values, all the
others correspond to the manufacturer's specifications.

Temp.

(°C)

0
5

10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95

13.423
13.207
13.003
12.810
12.627

12.454

12.289
12.133
11.984

11.841

11.705
11.574
11.449

-
-
-
-
-
-
-

9.464

9.392

9.331

9.277
9.228

9.184

9.144
9.107
9.076
9.046
9.018
8.985

8.962

8.941

8.921

8.902

8.885

8.867

8.850
8.833

6.984
6.950

6.922
6.900
6.880

6.865

6.853
6.843
6.837
6.834
6.833
6.834
6.836

6.840

6.845

6.852

6.859

6.867

6.877
6.886

4.010
4.004

4.001
4.001

4.003

4.008

4.015
4.025
4.036
4.049
4.064
4.075

4.091

4.108

4.126

4.145

4.164

4.185

4.205
4.227

-

1.668
1.670
1.672
1.676

1.680

1.685

1.691

1.697
1.704
1.712
1.715
1.723

1.732

1.743

1.754

1.766

1.778

1.792
1.806

9.27
9.18
9.13
9.08
9.04

9.00

8.96
8.93
8.90
8.87
8.84

8.81

8.79
8.76
8.74
8.73

8.71

8.70
8.68
8.67

7.11

7.08
7.06
7.04
7.02

7.00

6.99
6.98
6.98
6.97
6.97
6.97
6.97
6.98
6.98
6.99
7.00
7.00

7.01

7.02

3.99
3.99
3.99
3.99
3.99

4.00

4.00

4.01

4.02
4.03

4.04
4.06
4.07
4.09

4.11

4.13
4.15
4.18
4.20
4.23

NIST13

pH 13

NIST9

pH 9

NIST7

pH 7

NIST4

pH 4

NIST1

pH 1

Met9

pH 9.00

Met7

pH 7.00

Met4

pH 4.00

NIST (DIN 19266, 2000)

Metrohm

Advertising