3 dosing – Metrohm 916 Ti-Touch User Manual

Page 270

Advertising
background image

27.3 Dosing

■■■■■■■■■■■■■■■■■■■■■■

254

■■■■■■■■

916 Ti-Touch

Measuring mode

Selection of the measuring mode. Only those measuring modes are dis-
played which are advisable for the selected sensor.

Selection

pH | U | Ipol | Upol | T

pH
potentiometric pH measurement
U
potentiometric voltage measurement
Ipol
voltametric measurement with selectable polarization current
Upol
amperometric measurement with selectable polarization voltage
T
Temperature measurement

Temperature

Temperature entered manually. If a temperature sensor is connected then
the temperature will be measured continuously. This value is used for tem-
perature correction in pH measurements.

Input range

–20.0 - 150.0 °C

Default value

25.0 °C

[Info sensor]

Display information on the sensor.

27.3

Dosing

Manual control

Dosing

The following dosing functions are available:

Preparing the exchange or dosing unit (see Chapter 27.3.3, page 261)

Emptying the dosing unit (see Chapter 27.3.4, page 262)

Filling the dosing cylinder of the exchange unit/dosing unit (see Chap-
ter 27.3.5, page 262)

Dosing a specified volume (see Chapter 27.3.2, page 258)

Dosing continuously (see Chapter 27.3.1, page 257)

Selecting the dosing device
If several dosing devices are connected to a control device, then the dia-
log for selecting a dosing device will be opened.

Advertising